Description
Fluke 725 Multifunction Process Calibrator adalah kalibrator proses industri profesional yang dirancang untuk memberikan akurasi tinggi dan fleksibilitas maksimal dalam pengujian serta kalibrasi berbagai instrumen proses. Alat ini sangat cocok digunakan oleh teknisi instrumentasi, insinyur pemeliharaan, dan kalibrator profesional di industri minyak & gas, pembangkit listrik, manufaktur, serta otomasi industri.
FITUR Fluke 725
- Kalibrasi pemancar suhu dan tekanan
- Ukur/sumber/simulasikan parameter loop mA
- Ukur dan sumber mA, Volt DC, frekuensi, dan resistansi
- Catu daya loop 24 V bawaan
- Simpan dan ingat pengaturan kalibrasi yang sering digunakan
SPESIFIKASI Fluke 725
| Measurement Accuracy | ||
| Voltage DC | 30.000 V | 0.02% + 2 counts (upper display) |
| 30.000 V | 0.02% + 2 counts (lower display) | |
| 100.00 mV | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| Current DC | 24.000 mA | 0.02% + 2 counts |
| Resistance | 0.0 to 400.0 Ω | 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire) |
| 401 to 1500 Ω | 0.5 Ω (4-wire), 1 Ω (2- and 3-wire) | |
| 1500 to 3200 Ω | 1 Ω (4-wire), 1.5 Ω (2- and 3-wire) | |
| Frequency | 2.0 to 1000.0 CPM | 0.05% + 1 count |
| 1.0 to 1100.0 Hz | 0.05% + 1 count | |
| 1.00 to 10.00 kHz | 0.05% + 1 count | |
| Sensitivity | 1 V peak-to-peak-minimum | |
| Pressure | Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. | |
| Source Accuracy | ||
| Voltage DC | 100.00 mV | 0.02% + 2 counts |
| 10.000 V | 0.02% + 2 counts | |
| -10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
| Current DC | 24.000 mA (source) | 0.02% + 2 counts |
| 24.000 mA (simulate) | 0.02% + 2 counts | |
| Resistance | 15.0 to 400.0 Ω | 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA) |
| 401 to 1500 Ω | 0.5 Ω (excitation current 0.05 to 0.8 mA) | |
| 1500 to 3200 Ω | 1 Ω (excitation current 0.05 to 0.4 mA) | |
| Frequency | 2.0 to 1000.0 CPM | 0.05% |
| 1.0 to 1100.0 Hz | 0.05% | |
| 1.00 to 10.00 kHz | 0.25% | |
| Waveform | 5 V p-p squarewave, -0.1 V offset | |
| RTDs and Thermocouples | ||
| Measure accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.8°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Source accuracy | NI-120 | 0.2°C |
| PT-100 (385) | 0.33°C | |
| PT-100 (393) | 0.3°C | |
| PT-100 (JIS) | 0.3°C | |
| PT-200 (385) | 0.2°C | |
| PT-500 (385) | 0.3°C | |
| PT-1000 (385) | 0.2°C | |
| Resolution | 0.1°C | |
| Accuracy stated for 4-wire measurement. | ||
| J | 0.7°C | |
| K | 0.8°C | |
| T | 0.8°C | |
| E | 0.7°C | |
| R | 1.4°C | |
| S | 1.5°C | |
| B | 1.4°C | |
| L | 0.7°C | |
| U | 0.75°C | |
| N | 0.9°C | |
| Resolution | J, K, T, E, L, N, U: 0.1°C, B, R, S: 1°C | |
| XK | 0.6°C | |
| BP | 1.2°C | |
| Specifications | ||
| Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature |
| Ramps | Slow ramp, Fast ramp, 25% step-ramp | |
| Loop power function | Voltage | 24 V |
| Accuracy | 10% | |
| Maximum current | 22 mA, short circuit protected | |
| Step functions | Source functions | Voltage, current, resistance, frequency, temperature |
| Steps | 25% of range, 100% of range | |
| Environmental Specifications | ||
| Operating temperature | -10°C to +55°C | |
| Storage temperature | -20°C to 71°C | |
| Operating altitude | 3000 m | |
| Safety Specifications | ||
| Agency approvals | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | |
| Mechanical and General Specifications | ||
| Size | 130 x 236 x 61 mm | |
| Weight | 0.65 kg | |
| Batteries | 4 AA alkaline batteries | |
| Battery replacement | Separate battery compartment, accessable without breaking calibration seal | |
| Side port connections | Pressure module connector, also used for remote real-time programming | |